首页 | 本学科首页   官方微博 | 高级检索  
     


Measurement of left ventricular ejection fraction by videodensitometric analysis of digital subtraction angiograms
Authors:Jonathan Tobis  Orhan Nalcioglu  Anthony Seibert  W.David Johnston  Walter L. Henry
Affiliation:From the Division of Cardiology and Department of Radiological Sciences, University of California, Irvine, Orange, California USA
Abstract:
Left ventricular ejection fraction (LVEF) was calculated from 25 first-pass digital subtraction angiograms using a densitometric analysis. Digital subtraction angiograms are obtained in a computerized format; therefore, they can be readily analyzed with computer software to measure the density of the iodine signal within the image. The video signals from the image intensifier were logarithmically amplified so that there was a linear correlation between the video signal intensity and the depth of the iodine contrast material represented by that video signal. LVEF was also calculated by the area-length method from the same digital subtraction angiograms. There was close correlation between these two techniques (r = 0.94, standard error of the estimate = 5.04%). The videodensitometric EF technique is simple to perform, it correlates well with the standard area-length method, and is not dependent on geometric assumptions of LV geometry.
Keywords:Address for reprints: Jonathan Tobis   MD   Division of Cardiology   University of California   Irvine Medical Center   Route 81   101 City Drive South   Orange   California 92668.
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏