首页 | 本学科首页   官方微博 | 高级检索  
检索        

Effect of Different Blood Cycle Modes on Mass Transfer of Artificial Kidney
作者姓名:LIU  Tao  He  Li-qun  DING  Wei-ping  Zhao  Gang  GAO  Da-yong
作者单位:[1]Department of Thermal Science & Energy Engineering,USTC,Hefei 230027,China [2]Department of Mechanical Engineering, University of Kentucky, Lexington, KY40506, USA
基金项目:The project supported by the National Natural Science Foundation of China (5016016) and Natural Science Foundation of Anhui Province(03043717)
摘    要:INTRODUCTION Hollow fiber dialyzer has been widely applied to hemodialysis and chemicaldialysis process, and it is also called artificial kidney in medical science. There are ofgreat importance to improve the clinical dialysis efficiency by studying the masstransfer performance under different blood cycle modes in artificial kidney. Over the past decades, most models have been developed to describe transportphenomena in artificial kidney, the classical models are double-porous media model…

关 键 词:血液循环  循环方式  人工肾  血行转移  生物技术

Effect of Different Blood Cycle Modes on Mass Transfer of Artificial Kidney
LIU Tao He Li-qun DING Wei-ping Zhao Gang GAO Da-yong.Effect of Different Blood Cycle Modes on Mass Transfer of Artificial Kidney[J].Chinese Journal of Biomedical Engineering,2005,14(2):65-71.
Authors:LIU Tao  He Li-qun  DING Wei-ping  Zhao Gang  GAO Da-yong
Abstract:Based on the one-dimensional and unsteady-state Krogh model, this paper investigates mass transfer of artificial kidney under three blood cycle modes during the course of hemodialysis. The variations of the permeable solute clearance with increasing time and the dialysis time with increasing blood flux are simulated in detail ,and then one optimal blood cycle mode is acquired. The results are very important to improve the clinical dialysis efficiency of artificial kidney.
Keywords:Hollow fiber  Artificial kidney  Mass transfer  Unsteady state  Blood cycle mode