Development of a wearable multi-frequency impedance cardiography device
Abstract:Abstract

Cardiovascular diseases as well as pulmonary oedema can be early diagnosed using vital signs and thoracic bio-impedance. By recording the electrocardiogram (ECG) and the impedance cardiogram (ICG), vital parameters are captured continuously. The aim of this study is the continuous monitoring of ECG and multi-frequency ICG by a mobile system. A mobile measuring system, based on ‘low-power’ ECG, ICG and an included radio transmission is described. Due to the high component integration, a board size of only 6.5?cm?×?5?cm could be realized. The measured data can be transmitted via Bluetooth and visualized on a portable monitor. By using energy-efficient hardware, the system can operate for up to 18?hs with a 3?V battery, continuously sending data via Bluetooth. Longer operating times can be realized by decreased transfer rates. The relative error of the impedance measurement was less than 1%. The ECG and ICG measurements allow an approximate calculation of the heart stroke volume. The ECG and the measured impedance showed a high correlation to commercial devices (r?=?0.83, p?<?0.05). In addition to commercial devices, the developed system allows a multi-frequency measurement of the thoracic impedance between 5–150?kHz.
Keywords:Impedance cardiography  multi-frequency  personal healthcare  portable