首页 | 本学科首页   官方微博 | 高级检索  
     


Canadian Registry of Implantable Electronic Device Outcomes: Surveillance of High-Voltage Leads
Authors:Ratika Parkash  Bernard Thibault  Francois Philippon  Raymond Yee  Elizabeth Stephenson  Jeff Healey  Andrew Krahn  Derek Exner  Christopher Simpson  Eugene Crystal  Pablo Nery  Vidal Essebag  Laurence Sterns  Anthony Tang  George Wells
Affiliation:1. Department of Medicine, Division of Cardiology, QEII Health Sciences Center, Halifax, Nova Scotia, Canada;2. Division of Cardiology, Montréal Heart Institute, Montréal, Québec, Canada;3. Division of Cardiology, Université de Laval, Québec City, Québec, Canada;4. Division of Cardiology, London Health Sciences Center, London, Ontario, Canada;5. Division of Cardiology, Hospital for Sick Children, Toronto, Ontario, Canada;6. Division of Cardiology, Hamilton Health Sciences Center, Hamilton, Ontario, Canada;g. Division of Cardiology, University of British Columbia, Vancouver, British Columbia, Canada;h. Division of Cardiology, Libin Cardiovascular Institute, Calgary, Alberta, Canada;i. Division of Cardiology, Kingston General Hospital, Kingston, Ontario, Canada;j. Division of Cardiology, Sunnybrook Health Sciences Center, Toronto, Ontario, Canada;k. Division of Cardiology, University of Ottawa Heart Institute, Ottawa, Ontario, Canada;l. Divisions of Cardiology, McGill University and Hôpital de Sacre Coeur, Montréal, Québec, Canada;m. Division of Cardiology, Royal Jubilee Hospital, Victoria, British Columbia, Canada;n. Cardiovascular Research Methods Center, University of Ottawa Heart Institute, Ottawa, Ontario, Canada
Abstract:

Background

Cardiac implantable electrical devices (CIEDs) are subject to advisories and complications that can result in morbidity and mortality for patients; there is currently no system in Canada to track these.

Methods

This was a multicenter, prospective cohort study conducted at 5 centers to determine feasibility. Patients with a de novo high-voltage (HV) lead implantation were included and followed for a minimum of 1 year.

Results

There were 611 leads enrolled into the registry over 18 months. The mean age was 62.4 ± 12.8 years; 144 (23.6%) women were enrolled. The indication for lead implantation was for primary prevention in 65.5%. There were 497 (82.1%) de novo devices (single chamber: 54.5%, dual chamber: 20.5%, cardiac resynchronization therapy [CRT] 25.0%); the remainder of the procedures was a system revision for either upgrade (8.1%) or lead revision (9.8%). The lead revision rate at 1 year was 3.4%, with the primary reason being lead dislodgements. Mortality rate was 3.8% at 1 year. The rate of any device-related complication was 2.0% at 30 days, with the highest rate in CRT implants (4.9%, P = 0.0105). At 1 year, the complication rate was 4.5%, with no significant difference among device types.

Conclusions

This study demonstrates that device surveillance is feasible and highlights (1) the need for CIED surveillance to track device-related complications, (2) the scope of this should be larger, and (3) mandatory participation should be considered. This system could predict CIEDs that may be susceptible to higher than usual rates of failure, mitigating adverse outcomes in patients.
Keywords:Corresponding author: Dr Ratika Parkash   QEII Health Sciences Centre   HI Site   796 Summer Street   Room 2501D   Halifax   Nova Scotia B3H 3A7   Canada. Tel.: +1-902-473-4474. Fax: +1-902-473-3158.
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号