Irregular Pacemaker Tachycardia in a Patient with WPW Syndrome and an A-V Universal Pacemaker |
| |
Authors: | MICHAEL WEHR CLAUS-G. SCHMITT J. HERMANN PETER |
| |
Affiliation: | Department of Cardiology, Center for Internal Medicine, Philipps University, Marburg, Federal Republic of Germany;Center for Operative Medicine I, Philipps University, Marburg, Federal Republic of Germany |
| |
Abstract: | ![]() We describe a case of an irregular pacemaker circus movement tachycardia in a patient with the Wolff-Parkinson-White syndrome and a normally functioning A-V universal (DDD) pacemaker (Cordis Sequicor 233 D). The mechanism of the artificial circus movement tachycardia, which uses the pacemaker as the anterograde limb and a septally located accessory atrioventricular pathway as the retrograde limb, is discussed. |
| |
Keywords: | A-V universal pacing pacemaker tachycardia WPW syndrome Cordis Sequicor 233 D pacemaker |
|
|