首页 | 本学科首页   官方微博 | 高级检索  
     


Looking for a Learning Curve in EVAR Based on the Zenith Stent Graft
Authors:Manfred Kalteis  Peter Benedikt  Florian Huber  Florian Haller  Manfred Kastner  Herbert Lugmayr
Affiliation:1.Department of Cardiac, Vascular and Thoracic Surgery, Klinikum Wels-Grieskirchen, Wels, Austria;2.Department of Interventional Radiology, Klinikum Wels-Grieskirchen, Wels, Austria
Abstract:The purpose of this study was to report the learning curve of endovascular aneurysm repair (EVAR) based on the Zenith stent graft (Cook Medical Inc., Bloomington, IN). In the last 9 years, 101 patients were treated with a Zenith stent graft. To display the learning curve, a cumulative sum (CUSUM) failure analysis curve of the 30-day technical success rate was calculated. For detailed analysis, our EVAR patient cohort was chronologically divided into three groups. Technical and clinical results, basic patient parameters, and procedural data were compared. The CUSUM graph indicated an initial sharp rise within the first 35 cases and a plateau thereafter. The 30-day technical success rate significantly increased from the first to the second group (83 vs. 100%; p = 0.019), as did the primary technical success rate (66 vs. 97%; p = 0.001). EVAR based on the Zenith stent graft required ∼35 cases to reach a stably high rate of short-term technical success.
Keywords:abdominal aortic aneurysm   endovascular repair   endovascular procedure   EVAR   learning curve   Zenith stent graft   early results
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号