Sternal closure with titanium plate fixation--a paradigm shift in preventing mediastinitis |
| |
Authors: | Raman Jai Song David H Bolotin Gil Jeevanandam Valluvan |
| |
Affiliation: | Department of Surgery, Section of Plastic Surgery, University of Chicago, MC 5040, 5841 South Maryland Ave, Chicago, IL 60637, USA. jraman1@uchicago.edu |
| |
Abstract: | Sternal instability predisposes to post-operative mediastinitis. Biomechanical studies have shown the superiority of rigid plate fixation over wire circlage in sternal healing. We studied rigid plate fixation for sternotomies in high-risk patients. High-risk patients were identified as those having three or more historically established risk factors for post-operative mediastinitis, such as COPD, re-operative surgery, renal failure, diabetes, steroid use, obesity, existing infection, and immunosuppression. Three hundred and twenty high-risk patients had prophylactic rigid plate fixation (Group S) between July 2000 and Jan 2005. The control group (Group C) comprised 215 patients with similar risk profiles that were not plated during 2000 and 2001. Average age, male-female ratio, risk factors and type of procedures were similar in both groups. Follow up ranged from 4 to 200 weeks. There were 12 peri-operative deaths (3.75%) in group S and 8.6% (18 patients) in group C. There were no instances of deep mediastinitis in group S. Group C had mediastinitis in 28 (13%, P<0.05), requiring high dose antibiotics and plastic surgical intervention. Sternal fixation with titanium plates is an effective way of ensuring sternal immobility thereby reducing the substrate for bony infections. Application of this technique in high-risk patients prevents mediastinitis. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|