声触诊组织量化和弹性成像技术评价乳腺良、恶性肿瘤
引用本文:张渊,江泉,陈剑,张云霄,何竹,顾晓鸣.声触诊组织量化和弹性成像技术评价乳腺良、恶性肿瘤[J].中国医学影像技术,2013,29(2):206-209.
作者姓名:张渊  江泉  陈剑  张云霄  何竹  顾晓鸣
作者单位:上海市浦东新区人民医院超声科,上海 201200;上海市浦东新区人民医院超声科,上海 201200;上海市浦东新区人民医院超声科,上海 201200;上海市浦东新区人民医院超声科,上海 201200;上海市浦东新区人民医院超声科,上海 201200;上海市浦东新区人民医院医学统计室,上海 201200
基金项目:上海市浦东新区卫生局卫生科技发展专项基金(PW2011A-15)、上海市浦东新区卫生系统优秀青年医学人才培养计划(PWRq2010-09)、上海市卫生局科研项目(20114Y165)、浦东新区卫生系统领先人才培养计划(PWR12012-02)。
摘    要:目的探讨声触诊组织量化(VTQ)和弹性成像技术诊断乳腺良、恶性肿瘤的价值。方法对86例乳腺病变患者术前分别进行常规超声、弹性超声及VTQ检查,对各病灶进行弹性成像评分和VTQ速度分组,并与病理结果对照。结果乳腺良、恶性肿瘤弹性超声评分差异有统计学意义(P<0.05),良性肿瘤中弹性超声评分为1~3分者占为68.09%(32/47),恶性肿瘤中弹性超声4~5分者占82.05%(32/39)。乳腺良、恶性肿瘤的VTQ速度值差异亦有统计学意义(P<0.05)。常规超声、弹性超声、VTQ技术联合应用的诊断准确率达91.86%。结论联合应用常规超声、弹性成像、VTQ技术,可显著提高诊断乳腺良、恶性肿瘤的准确率。

关 键 词:乳腺肿瘤  超声检查  弹性成像技术  声触诊组织量化
收稿时间:7/3/2012 12:00:00 AM
修稿时间:2012/12/5 0:00:00

Virtual touch tissue quantification and ultrasonic elastography in differential diagnosis of benign and malignant breast tumors
ZHANG Yuan,JIANG Quan,CHEN Jian,ZHANG Yun-xiao,HE Zhu and GU Xiao-ming.Virtual touch tissue quantification and ultrasonic elastography in differential diagnosis of benign and malignant breast tumors[J].Chinese Journal of Medical Imaging Technology,2013,29(2):206-209.
Authors:ZHANG Yuan  JIANG Quan  CHEN Jian  ZHANG Yun-xiao  HE Zhu and GU Xiao-ming
Institution:Department of Ultrasound, Pudong New Area People's Hospital, Shanghai 201200, China;Department of Ultrasound, Pudong New Area People's Hospital, Shanghai 201200, China;Department of Ultrasound, Pudong New Area People's Hospital, Shanghai 201200, China;Department of Ultrasound, Pudong New Area People's Hospital, Shanghai 201200, China;Department of Ultrasound, Pudong New Area People's Hospital, Shanghai 201200, China;Department of Medical Statistics, Pudong New Area People's Hospital, Shanghai 201200, China
Abstract:Objective To investigate the diagnostic value of virtual touch tissue quantification (VTQ) and ultrasonic elastography for benign and malignant breast tumors. Methods Conventional ultrasonography, VTQ and ultrasonic elastography were used to examine 86 patients with breast lesions before operation, and ultrasonic elastography score and VTQ velocity classification were done for every lesion. The diagnostic results were contrasted with pathology after operation. Results There was significant difference in detectable rate of ultrasonic elastography score between benign and malignant groups (P<0.05). There were 68.09% (32/47) benign breast tumor scored 1 to 3, whereas 82.05% (32/39) malignant breast tumors scored 4 to 5. There was significant difference in VTQ velocity classification between benign and malignant groups (P<0.05). The diagnostic accuracy of combination of three methods was 91.86%. Conclusion Combined application of conventional ultrasound, VTQ and ultrasonic elastography can improve the diagnostic accuracy of benign and malignant breast tumors.
Keywords:Breast neoplasms  Ultrasonography  Elasticity imaging technique  Virtual touch tissue quantification