Flow and peak velocity measurements in patients with aortic valve stenosis using phase contrast MR accelerated with k-t BLAST |
| |
Authors: | Thunberg Per Emilsson Kent Rask Peter Kähäri Anders |
| |
Affiliation: | 1. Department of Medical Physics, Örebro University Hospital, S-70185 Örebro, Sweden;2. Department of Clinical Physiology, Örebro University Hospital, Sweden;3. Department of Radiology, Örebro University Hospital, Sweden |
| |
Abstract: |
ObjectiveTo investigate the accuracy of velocity measurements in patients with aortic valve stenosis using phase contrast (PC) imaging accelerated with SENSE (Sensitivity Encoding) and k-t BLAST (Broad-use Linear Acquisition Speed-up Technique).MethodsAccelerated quantitative breath hold PC measurements, using SENSE and k-t BLAST, were performed in twelve patients whose aortic valve stenosis had been initially diagnosed using echocardiography. Stroke volume (SV) and peak velocity measurements were performed on each subject in three adjacent slices using both accelerating methods.ResultsThe peak velocities measured with PC MRI using SENSE were −8.0 ± 9.5% lower (p < 0.01) compared to the peak velocities measured with k-t BLAST and the correlation was r = 0.83. The stroke volumes when using SENSE were slightly higher 0.4 ± 17.1 ml compared to the SV obtained using k-t BLAST but the difference was not significant (p > 0.05).ConclusionsIn this study higher peak velocities were measured in patients with aortic stenosis when combining k-t BLAST with PC MRI compared to PC MRI using SENSE. A probable explanation of this difference is the higher temporal resolution achieved in the k-t BLAST measurement. There was, however, no significant difference between calculated SV based on PC MRI using SENSE and k-t BLAST, respectively. |
| |
Keywords: | Aortic valve Stenosis k-t BLAST SENSE Phase contrast imaging |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|