Experimental evaluation of a new transcatheter vascular embolization device in the swine model
Authors:Salamat Mehrdad  Brown Phillip R  Magee Carolyn A  Reyes Diane K  Peters David N  Venbrux Anthony C
Affiliation:Division of Pediatric Cardiology, The Johns Hopkins Medical Institutions, Baltimore, MD, USA.
Abstract:
PURPOSE: To evaluate a new transcatheter device suitable for arterial embolization in an animal model. MATERIALS AND METHODS: A new prototype self-expanding braided embolic device (Embolizor), consisting of nitinol wire strands fixed at either end with platinum-iridium bands and covered with a film of polyethylene, was deployed through 5-F diagnostic catheters into renal artery branches in five swine. Standard stainless-steel spring coils were deployed in other renal branches and served as controls. The animals underwent follow-up selective renal arteriography to determine presence or absence of vessel recanalization or device migration 15-23 days after device deployment. Histopathologic evaluation of target vessels and peripheral renal parenchyma was also performed. RESULTS: Ten Embolizors and 10 stainless-steel coils were deployed in arteries ranging in size from 1.8 to 3.0 mm in diameter. The Embolizor was easily and precisely deployed. Angiographic evidence of vascular occlusion in the Embolizor group was noted within 30 seconds in eight device deployments and within 5 minutes in two. No early or delayed device migration was noted on follow-up arteriography. In the control group, seven of 10 previously occluded arteries were recanalized. No recanalization was noted in the Embolizor group. Light microscopy revealed evidence of infarction in all specimens examined. Whereas three specimens in the Embolizor group contained occasional giant cells, there were numerous multinucleated giant cells present within the interstices of all control spring coils. CONCLUSION: The Embolizor was easily, precisely, and successfully deployed through standard selective diagnostic angiographic catheters. Short-term follow-up demonstrated no recanalization or migration of the device. The Embolizor was shown on histopathologic analysis to have no significant foreign body reaction.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号