首页 | 本学科首页   官方微博 | 高级检索  
     


Coronary sinus cannulation with a steerable catheter during biventricular device implantation
Authors:Lingwei Wang  Shiwen Yuan  Rasmus Borgquist  Carl-Johan Höijer  Johan Brandt
Affiliation:Department of Arrhythmias, Skane University Hospital, Lund University,Lund, Sweden
Abstract:
Objectives. To determine whether a steerable catheter with electrogram guidance (CS-assist group) could facilitate access to the coronary sinus (CS) during cardiac resynchronization therapy (CRT) implantation. Design. Consecutive patients who underwent CRT implantation were recruited prospectively into the CS-assist group (n = 81) and compared with those using conventional techniques without an electrogram guidance (conventional group, n = 101). Results. The CS cannulation success rate was clearly greater in the CS-assist group (100%) than that in the conventional group (95%, p < 0.05), with significantly shorter mean procedure time (52.6 ± 20.6 min vs. 73.2 ± 40.9 min, p < 0.01) and fluoroscopy time (3.6 ± 3.2 min vs. 14.2 ± 20.4 min, p < 0.01). In the five CS cannulation failure cases, mean procedure time (144.0 ± 37.0 min) and fluoroscopy time (57.8 ± 24.8 min) were significantly longer than those in the other patients (61.2 ± 32.3 and 8.2 ± 13.6 min, respectively, n = 177, both p < 0.01). Conclusions. Using the steerable catheter with real-time electrogram guidance, location of and access to the CS is more rapid and successful, which may improve the success of the CRT implantation and may give significant time savings.
Keywords:cardiac resynchronization therapy  coronary sinus cannulation  electrogram guidance  heart failure