Affiliation:(1) Department of Diagnostic Radiology (Division of Nuclear Medicine), University Hospital Leiden, P.O. Box 9600, 2300 RC Leiden, The Netherlands;(2) Division of Radiation Protection, Ministry of Housing, City Planning and Environment, The Hague, The Netherlands;(3) Department of Clinical Oncology, University Hospital, Leiden, The Netherlands
Abstract:The construction of a breathing lung phantom that can be used to measure the amount of radioactive gas in the lungs as well as to determine the absorbed dose is described. For a lung ventilation study that consists of 6 views of 300 kents each, an effective dose equivalent of 50 mgrSv was calculated. The phantom is also suitable for comparison of different generator systems.
Keywords:81mKr lung ventilation   81mKr effective dose equivalent  Lung phantom