摘    要:
目的 观察UF-100尿沉渣自动分析仪CHECK液前向散射光强度(FSC)失控对尿液样本红细胞FSC和P70FSC指标的影响.方法 使用UF-100 CHECK液对UF-100进行校准,并于检测前后分别测试70份尿液样本和50份用生理盐水稀释的全血质控物样本并记录校准前37 d和校准后23 d UF-100质量控制结果,使用SPSS11.0对所得数据进行统计学分析.结果 UF-100 FSC失控可在五项定量指标在控情况下出现;校准前后红细胞光学指标FSC、P70FSC和FSC-SDW存在统计学差异;对尿液样本的线性分析可见校准后FSC和P70FSC较校准前分别出现 10.85和 21.27ch漂移.结论 UF-100 CHECK液FSC减低失控提示系统所测定尿液样本红细胞FSC和P70FSC测定值会出现漂移;UF-100质量控制前向散射光强度的减低,势必造成UF-100系统对尿液颗粒的鉴别能力降低;FSC质量控制是UF-100在肾性血尿的鉴别诊断中应用的前提.

关 键 词:UF-100尿沉渣分析仪  前向散射光强度  质量控制  血尿
文章编号:1008-7044(2007)05-0405-02
修稿时间:2007-05-31

The influence of losing control of the UF-100''''s CHECK liquid FSC index on urine sediment''''s RBC optical index
LI Yan-xin,ZHONG Ya-ling. The influence of losing control of the UF-100''''s CHECK liquid FSC index on urine sediment''''s RBC optical index[J]. Journal of Huaihai Medicine, 2007, 25(5): 405-406
Authors:LI Yan-xin  ZHONG Ya-ling
Abstract:
Objective To observe the influence of losing control of CHECK liquid FSC index of UF-100 on urine RBC's FSC and P70FSC index.Mechods The UF-100 CHECK liquid was used to calibrate UF-100 system,and to measure 70 urine samples and 50 full blood quality control samples diluted by physiologic brine before and after calibration.The UF-100 quality control results 37 days before calibration and 23 days after calibration were noted down and analyzed with SPSS 11.0.Results FSC UF-100s' losing control could appear when the five quantitative index were under control.By the linear analysis of urine samples it was proved there was +10.85 and+21.27 ch drift respectively between after and before calibration in FSC and P70FSC index of red blood cell.Conclusion The UF-100 CHECK liquid FSC decrease and losing control indicate that urine RBC's FSC and P70FSC index would drift.The decrease of FSC of the UF-100 quality control will certainly reduce the discrimination ability of the UF-100 system.FSC quality control is necessary before UF-100 in discriminating diagnosis of renal blood urine.
Keywords:UF-100 urinary sediment analyzer   FSC   Quality control   Blood urine