Low flow veno-venous ECMO: an experimental study |
| |
Authors: | Calderón M Verdín R Galván J Gonzalez M Cárdenas H Campos R Vidrio H Amezcua J |
| |
Affiliation: | Department of Circulatory Support, "La Raza" Medical Center and Military School of Medicine, México City. |
| |
Abstract: | Clinical use of extracorporeal membrane oxygenation (ECMO) and carbon dioxide removal (ECCO 2R) have become well established techniques for the treatment of severe respiratory failure; however they require full cardiopulmonary bypass, representing major procedures with high morbidity. We theorized the possibility of an efficient low flow veno-venous extracorporeal membrane gas exchange method. Four mongrel 12 kg dogs were submitted to veno-venous extracorporeal membrane gas exchange via a jugular dialysis catheter using a low flow (10 ml/min) roller pump and a membrane oxygenator for a period of four hours. Respiratory rate was set at 4 breaths/min with a FiO 2 of 21% and ventilatory dead space was increased. Adequate gas exchange was obtained (pO 2139, pCO 224, Sat 99.4%), without major hemodynamic changes or hematuria. Our results demonstrate the feasibility of a low flow, less aggressive system. Further research should be considered. |
| |