Arthrosynovial cyst of the hip after total hip arthroplasty with a ceramic-on-ceramic interface |
| |
Authors: | Legaye J Lenfant P Delos M |
| |
Affiliation: | Service de Chirurgie Orthopédique et Traumatologique, Cliniques Universitaires U.C.L. de Mont-Godinne, 5530 Yvoir, Belgique. |
| |
Abstract: | An inguinal arthrosynovial cyst of the hip joint developed after total hip arthroplasty leading to compression of the femoral vein. This complication occurred after implantation of a non-cemented prosthesis with a ceramic-on-ceramic interface inserted in a polyethylene sandwich. Multiple foreign bodies composed of both ceramic and polyethylene particles were visualized microscopically. Later, rupture of the joint interface required implant replacement. This cyst was a precursor sign of a deteriorating prosthesis surface. |
| |