Transapical mitral valved stent implantation: Enhanced survival and decreased paravalvular leakages
Authors:Saskia Pokorny  Katharina Huenges  Telse Bähr  Jan Hinnerk Hansen  Gunther Fischer  Justus Gross  Michael Morlock  Jochen Cremer  Georg Lutter
Institution:1. Department of Cardiovascular Surgery, Christian-Albrechts-University of Kiel, School of Medicine, Arnold-Heller-Str. 3, D-24105 Kiel, Germany;2. Department of Pediatric Cardiology, Christian-Albrechts-University of Kiel, School of Medicine, Arnold-Heller-Str. 3, D-24105 Kiel, Germany;3. Institute of Biomechanics, TUHH Hamburg University of Technology, Denickestraße 15, D-21073 Hamburg, Germany
Abstract:

Background/Objectives

This study presents the recent results of transapical implantation of a new design of a mitral valved stent with up to three months follow-up.

Methods

A self-expanding re-designed mitral valved stent was implanted via transapical approach into the beating heart of eight pigs. Echocardiographic- and hemodynamic parameters were assessed before (n = 8), at 1 h (n = 8), at one month (n = 6), at two months (n = 2), and at three months (n = 1) after implantation and a cardiac CT was conducted.

Results

The stent was successfully deployed in all animals. Two animals died within the first month due to incorrect fixation force. Echocardiographic evaluation showed low gradients (3.9 ± 1.4 mm Hg and 1.9 ± 0.8 mm Hg across the valved stent and aortic valve) and a normal mitral annular plane systolic excursion (1.1 ± 0.2 cm) after one month. No paravalvular leakages (PVL) were detected after 1 h. The pulmonary artery pressure did not increase after valved stent implantation (p ≥ 0.106). The pulmonary capillary wedge pressure (PCWP) slightly increased to 16 ± 3 mm Hg after one month (p = 0.033). The left ventricular end-diastolic pressure was mildly elevated (15.8 ± 8.6 mm Hg) after one month.

Conclusions

Secure deployment and correct position of the valved stents were reproducibly achieved in the off-pump implantation procedure. No paravalvular leakages after 1 h as well as low gradients, few stent fractures and a normal longitudinal function after one month were achieved with this newly developed and well-aligned prototype. However, a number of challenges have been identified during this study and potential for improvement has been identified.
Keywords:Off-pump  Transapical  Valved stent  Mitral valve  Stent design  Transcatheter