某实验室外照射个人剂量盲样比对结果分析
引用本文:蔡菁婷. 某实验室外照射个人剂量盲样比对结果分析[J]. 环境与职业医学, 2014, 0(5): 400-402
作者姓名:蔡菁婷
作者单位:闸北区疾病预防控制中心理化检验科,上海200072
摘    要:[目的]通过个人剂量盲样比对检验参比实验室个人剂量监测系统整体技术水平。[方法]按照《全国放射工作人员个人剂量监测系统比对方案》,某实验室2010年至2012年连续3次参加由中国疾病预防控制中心辐射防护与核医学安全所组织的全国外照射个人剂量监测盲样比对。[结果]2010年比对实际测量值与约定真值的相对偏差在±18%以内;2011年和2012年单项性能和综合性能均在±10%以内,3次盲样比对结果均符合国家相关标准要求。[结论]该实验室个人剂量监测系统整体技术水平符合国家相关标准要求。

关 键 词:个人剂量  盲样比对  质量控制  结果分析

Blind Sample Measurements of Personal External Exposure Dose in a Laboratory
CAI Jing-ting. Blind Sample Measurements of Personal External Exposure Dose in a Laboratory[J]. Journal of Environmental & Occupational Medicine, 2014, 0(5): 400-402
Authors:CAI Jing-ting
Affiliation:CAI Jing-ting (Physical and Chemical Test Department, Zhabei District Center for Disease Control and Prevention, Shanghai 200072, China)
Abstract:[ Objective ] To assess the overall technical level of personal dose monitoring system of a laboratory through random blind sample comparison of personal doses. [ Methods ] According to the National Comparison Protocol for Personal Dose Monitoring System for Radiation Emission Staff, blind samples of personal external exposure doses were collected from radiation emission workers of a laboratory attending three batches of the yearly blind sample examination organized by the National Institute for Radiological Protection, China Center for Disease Control and Prevention from 2010 to 2012. [ Results ] The relative deviation of the measured values from conventional true values lied within + 18% in 2010. Both single unit determination and comprehensive deiermination lied within + 10% in 2011 and 2012. The comparison results were qualified. [ Conclusion ] The overall technical level of personal dose monitoring system in the laboratory meets the requirements of relevent national standards.
Keywords:personal dose  blind sample comparison  quality control  analysis of results