一款新型糖化血红蛋白床旁检测系统线性评价
引用本文:王尉,郑华荣,周睿,汤小琨. 一款新型糖化血红蛋白床旁检测系统线性评价[J]. 标记免疫分析与临床, 2017, 24(4). DOI: 10.11748/bjmy.issn.1006-1703.2017.04.024
作者姓名:王尉  郑华荣  周睿  汤小琨
作者单位:首都医科大学附属北京地坛医院输血科,北京,100015;中国计量科学研究院,北京,100013;首都医科大学附属北京朝阳医院检验科,北京,100020;天津医科大学临床医学院,天津,300070
基金项目:首都卫生发展科研专项项目《POCTHbA1c方法标准化的建立与应用评价及质谱检测HbA1c用于糖尿病分子分型方法评价》,国家临床重点专科建设项目,北京医院协会研究项目
摘    要:目的 对一款新型床旁检测(point-of-care testing,POCT)糖化血红蛋白(HbA1c)分析仪的线性进行评价,从而对这仪器及试剂进行初步评价.方法 根据临床和实验室标准化协会(CLSI) EP6-A文件,配制11个HbA1c浓度为等间距排列的试验样本.按随机顺序对它们进行测定,每个样本重复测定两次,对试验数据进行多项式回归分析.结果 测定结果未发现离群点.试验样本的不精密度变异系数(CVr)为2.4%.二次回归多项式非线性系数b2与0之间的差异无统计学意义(P=0.743),三次回归多项式非线性系数b2、b3与0之间的差异无统计学意义(P =0.869、0.847),最适多项式为一次多项式,最适方程为Y=0.971X +2.899(r =0.998).结论 该HbA1c检测系统在3.90%~13.45%范围内呈线性,该试剂盒申明的线性范围与本验证试验结果一致.

关 键 词:床旁检测  糖化血红蛋白  线性  评价方法

The Linearity Evaluation of a New Hemoglobin A1c Analyzer(HbA1c) for Point-of-Care Testing (POCT)
WANG Wei,ZHENG Hua-rong,ZHOU Rui,TANG Xiao-kun. The Linearity Evaluation of a New Hemoglobin A1c Analyzer(HbA1c) for Point-of-Care Testing (POCT)[J]. Labeled Immunoassays and Clinical Medicine, 2017, 24(4). DOI: 10.11748/bjmy.issn.1006-1703.2017.04.024
Authors:WANG Wei  ZHENG Hua-rong  ZHOU Rui  TANG Xiao-kun
Abstract:Objective To evaluate the linearity of a new hemoglobin Alc analyzer for point-of-care testing,as well as its instrument and test kit.Methods According to the EP6-A document,described by Clinical and Laboratory Standards Institute (CLSI),11 HbA1c concentrations of equal spaced were prepared as test samples,which were measured in duplicate in random order.Then the data were analyzed with polynomial regression analysis.Results No outlier was found in all the results.The coefficient variation of imprecision (CVr) is 2.4%.There was no statistically significant difference between zero and nonlinear coefficient b2 、b3 in quadratic regression polynomial (P =0.869,0.847),and the linear equation was optimal.The regression equation of HbA1c was Y =0.971X + 2.899 (r =0.998).Conclusion The HbA1c test kit was linear in the range of 3.90% ~ 13.45%,and the linear range stated in the kit was consistent with the results of this validation test.
Keywords:POCT  HbA1c  Linearity  Evaluation