Power density and external temperature of laser-treated root canals
Authors:Blankenau R J  Kelsey W P  Powell G L  Cavel W T  Anderson D M
Affiliation:Department of Operative Dentistry, Creighton University School of Dentistry, Omaha, NE.
Abstract:The purpose of this study was to determine the power and time parameters for an argon laser that would result in the removal of pulpal tissue without excessively elevating the external temperature of the root. External temperatures were measured by attaching thermistors to the surfaces of the teeth at the cemento-enamel junction (CEJ) areas and at the root apices. Results indicate that a 1-W power setting with a pulse duration of 0.1 sec and a 1 sec interval between pulses produced a mean temperature rise of 0.89 +/- 0.27 degrees C at the cemento-enamel junction area and a mean temperature increase of 2.04 +/- 0.47 degrees C at the apex. A 2-W power setting with a 0.1 sec pulse duration and a 1 sec interval between pulses resulted in a mean temperature increase of 1.58 +/- 0.45 degrees C at the CEJ and a mean temperature rise of 2.59 +/- 0.20 degrees C at the apex. Based upon the results of this study, it was concluded that an argon laser operating at 1 or 2 W of power with a 0.1 sec pulse duration could be used to remove pulpal tissue without creating an excessive increase in the external temperature of the tooth.