磁性附着体应用于口腔覆盖义齿修复的临床观察
引用本文:许培武.磁性附着体应用于口腔覆盖义齿修复的临床观察[J].广东牙病防治,2014(11):573-574.
作者姓名:许培武
作者单位:深圳市龙岗中心医院耳鼻咽喉科医院口腔科,广东深圳518116
摘    要:目的探讨磁性附着体在口腔修复中的临床应用效果。方法对42例患者应用磁性附着体行覆盖义齿修复,测量义齿修复前后的咀嚼效能,测量义齿的固位力。结果磁性附着体粘固即刻义齿的固位力为(886.25±0.41)g,咀嚼效能测定的光密度值为0.83±0.18,明显大于粘固前,差异有统计学意义(P〈0.05)。粘固后3周义齿的固位力为(890.31±0.31)g,咀嚼效能测定的光密度值为0.84±0.25,与粘固即刻比较差异无统计学意义(P〉0.05)。结论磁性附着体应用于口腔覆盖义齿修复,临床效果满意,值得推广应用。

关 键 词:磁性附着体  义齿固位  咀嚼效率

Clinical observation of oral magnetic attachment in overdentures
XU Pei-wu.Clinical observation of oral magnetic attachment in overdentures[J].Journal of Dental Prevention and Treatment,2014(11):573-574.
Authors:XU Pei-wu
Institution:XU Pei-wu.( Department of Stomatology, ENT Hospital of Shenzhen Longgang Central I-iospital, Shenzhen 518116, China)
Abstract:Objective To study the effect of clinical application of magnetic attachment in oral cavity repair. Meth- ods forty-two cases of patients were restored with magnetic attachment denture, measured masticatory function and den- ture retention force before and .after denture placement. Results Fixed force immediate denture cement was ( 886.25 ± 0.41 ) g, masticatory efficiency measured optical density value of 0.83 ± 0.18, significantly greater than before cementa- tion, the difference was statistically significant ( P 〈 0.05 ). After three weeks of denture cement retention force (890.31 ±0.31 ) g, masticatory efficiency ratio measured optical density value of 0.84 ± 0.25, compared with that of immediately after cementation, difference was not statistically significant ( P 〉 0.05). Conclusion Magnetic attachment used in over dentures show satistactory clinical results, and should be widely applied.
Keywords:Magnetic attachment  Denture retention  Mastieatory efficiency