Thrombogenicity of various endovascular stent types: an in vitro evaluation |
| |
Authors: | Tepe Gunnar Wendel Hans P Khorchidi Sascha Schmehl Jörg Wiskirchen Jakub Pusich Benjamin Claussen Claus D Duda Stephan H |
| |
Affiliation: | Department of Diagnostic Radiology, University of Tübingen, Hoppe-Seyler. Str. 3, 72076 Tübingen, Germany. gunnar.tepe@med.uni-tuebingen.de |
| |
Abstract: | PURPOSE: The aim of this study was to evaluate the thrombogenicity of different peripheral stent types in a standardized in vitro model with fresh human whole blood. MATERIALS AND METHODS: Different stents (N = 77; n = 7 of each of 11 types) were implanted in polyvinyl chloride tubing loops and filled with donor blood samples. After 120 minutes of blood circulation, the thrombin-antithrombin III complex (TAT) levels, beta-thromboglobulin (beta-TG) levels, and platelet counts were assessed. RESULTS: After 2 hours, significant differences were seen. TAT values (+/- SD) with the investigated stents were 31 micro g/mL +/- 20 (control, no stent), 328 micro g/mL +/- 206 (Saxx stent, peripheral medium CrNi31 L), 651 micro g/mL +/- 760 (Palmaz Corinthian Stent, 316 L stainless steel, electropolished), 1,609 micro g/mL +/- 1,264 (Palmaz Corinthian Stent, 316 L stainless steel, not electropolished), 810 micro g/mL +/- 578 (Palmaz Schatz long medium stent), 569 micro g/mL +/- 347 (Smart Nitinol stent), 1,037 micro g/mL +/- 577 (Megalink peripheral stent), 543 micro g/mL +/- 487 (peripheral stent, electropolished), 1,674 micro g/mL +/- 2,057 (peripheral stent, not electropolished), 3,128 micro g/mL +/- 1,812 (SelfX Nitinol stent, polished), 5,897 micro g/mL +/- 2,380 (SelfX Nitinol stent, unpolished), and 1,458 micro g/mL +/- 887 (bridge stent). The platelet count (x1,000/ micro L +/- SD) was 218 +/- 35 (control, no stent), 188 +/- 22 (Saxx stent), 187 +/- 20 (Palmaz Corinthian stent, electropolished), 135 +/- 37 (Palmaz Corinthian stent, not electropolished), 170 +/- 24 (Palmaz Schatz stent), 180 +/- 36 (Smart Nitinol stent), 159 +/- 26 (Megalink peripheral stent), 173 +/- 17 (peripheral stent, electropolished), 133 +/- 51 (peripheral stent, not electropolished), 123 +/- 37 (SelfX Nitinol stent, polished), 52 +/- 27 (SelfX Nitinol stent, unpolished), and 130 +/- 31 (bridge stent). CONCLUSION: This standardized study showed a wide range of platelet activation after stent implantation. Electropolishing clearly reduced the thrombogenicity of the stents. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|