Influence of the clearance on in-vitro tribology of large diameter metal-on-metal articulations pertaining to resurfacing hip implants |
| |
Authors: | Rieker Claude B Schön Rolf Konrad Reto Liebentritt Gernot Gnepf Patric Shen Ming Roberts Paul Grigoris Peter |
| |
Affiliation: | Tribology-113 955, Zimmer Europe GmbH, Sulzer Allee 8, CH 8404 Winterthur, Switzerland. claude.rieker@zimmer.com |
| |
Abstract: | ![]() Large-diameter metal-on-metal articulations may provide an opportunity for wear reduction in total hip implants because earlier studies have shown that the formation of a fluid film that completely separates the bearing surfaces is theoretically possible. In such a lubrication mode and under ideal conditions, there is theoretically no amount of wear. Studies have suggested that the two primary parameters controlling the lubrication mode are the diameter and the clearance of the articulation. The goal of the present study was to experimentally investigate the influence of these two parameters on the wear behavior of large-diameter metal-on-metal articulations pertaining to resurfacing hip implants. The results of this in vitro investigation showed that longer running-in periods and higher amounts of running-in wear were associated with larger clearances. |
| |