Performance evaluation of the PET sub-systems of two PET/CT imagers according to NEMA standard NU 2-2001
Authors:Strigl Markus  Martínez María-José  Zach Christian  Ziegler Sibylle I  Hahn Klaus
Affiliation:1. Klinikum der Universität München, Klinik und Poliklinik für Nuklearmedizin;2. Nuklearmedizinische Klinik, Klinikum rechts der Isar der Technischen Universität München;1. Department of Medical Radiation Engineering, Shahid Beheshti University, Tehran, Iran;2. Research Center for Nuclear Medicine, Shariati Hospital, Tehran University of Medical Sciences, Tehran, Iran;3. PET Center, Institute of Radiopharmaceutical Cancer Research, Helmholtz-Zentrum Dresden-Rossendorf, Dresden 01328, Germany;4. Department of Nuclear Medicine, University Hospital Carl Gustav Carus, Technische Universität Dresden, Dresden 01307, Germany;5. Departments of Radiology and Physics, University of British Columbia, Vancouver, BC, Canada;6. Department of Radiology and Radiological Science, Johns Hopkins University, Baltimore, MD, USA;1. Almetyevsk State Oil Institute, 2 Lenina St., Almetyevsk, Russia;2. South Ural State University, 76, Lenin Prospekt, Chelyabinsk, Russia
Abstract:
Performance and image quality of PET scanners are mainly determined by the choice of the detector material and the realised design of the scanner. The aim of this study was to measure the PET performance characteristics of two state-of-the-art PET/CT Systems. Methods: Performance measurements following the NEMA NU 2-2001 standard were performed on a GSO-based system and on a system with LSO detectors. The following parameters were determined: spatial resolution, scatter fraction, sensitivity and count rate behaviour. A body phantom was used to determine the image quality as well as the influence of differently performed attenuation corrections (low dose CT vs. 137Cs transmission source). Results: Both scanners have partly different, but altogether good performance characteristics. The differences can be explained mostly by different physical properties of the detectors and by a different design of the scanners including the implemented electronics. Conclusion: If the data are to be compared quantitatively, the modality of the attenuation correction has to be taken into account, especially in regions with reduced uptake or reduced density. Both scanners achieve a very good image quality rich in contrast within short acquisition times which allows easy detection of small lesions.
Keywords:PET/CT  PET  performance measurements  NEMA NU 2-2001  Gemini  Biograph  PET/CT  Leistungsdaten PET  NEMA NU 2-2001  Gemini  Biograph
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号