Atypical failure after cochlear implantation in children
Authors:S. Wartelle  M. BlanchardB. Thierry  M. ParodiI. Rouillon  E.N. GarabedianN. Loundon
Affiliation:Service ORL Hôpital Necker-Enfants malades, 149 rue de Sèvres, 75015 Paris, France
Abstract:
We report a case of intermittent dysfunction in a 10-years-old boy, implanted with MedEL® cochlear implant. Few weeks after the surgery the boy described short and intermittent episodes of implant dysfunction with rapid return to a normal function. No evidence for any electric or neural dysfunction was found. After few weeks, a clinical link was discovered to episodes of sneeze or nose blowing. Clinical and surgical implications are discussed.
Keywords:Children   Cochlear implant   Complication   Surgery   Failure   Skin flap