Lack of agreement between thermodilution and electrical velocimetry cardiac output measurements
Authors:Matthias Heringlake  Ulrich Handke  Thorsten Hanke  Frank Eberhardt  Jan Schumacher  Hartmut Gehring  Hermann Heinze
Institution:(1) Department of Anesthesiology, University of Luebeck, Ratzeburger Allee 160, 23538 Luebeck, Germany;(2) Department of Cardiac Surgery, University of Luebeck, Luebeck, Germany;(3) Department of Cardiology, University of Luebeck, Luebeck, Germany;(4) Department of Anaesthesia, St. Thomas Hospital, London, UK
Abstract:

Objective

The modified algorithm for the non-invasive determination of cardiac output (CO) by electrical bioimpedance—electrical velocimetry (EV®)—has been reported to give reliable results in comparison with echocardiography and pulmonary arterial thermodilution (PA-TD) in patients either before or after cardiac surgery. The present study was designed to determine whether EV®-CO measurements reflect intraindividual changes in CO during cardiac surgery.

Design

Prospective, observational study.

Setting

Operating room (OR) and intensive care unit (ICU) of a university hospital.

Patients

Twenty-nine patients undergoing elective cardiac surgery.

Interventions

None.

Measurements

CO was determined simultaneously by PA-TD and EV® after induction of anesthesia (t1) and 4.9?±?3.5?h after ICU admission (t2).

Results

TD-CO was 3.9?±?1.4 and 5.4?±?1.1 l/min at t1 and t2 (?p?®-CO was 4.3?±?1.1 and 4.9?±?1.5 l/min at t1 and t2 (?p?=?0.013). Bland–Altman analysis showed a bias of ?0.4 l/min and 0.4 l/min and a precision of 3.2 and 3.6 l/min (34.3% and 67.4%) at t1 and t2, respectively. Analysis of the individual pre- to postoperative changes in CO with both methods revealed bidirectional changes in n?=?12 patients and unidirectional changes with a difference greater than 50% and less than 50% in n?=?9 and n?=?8 patients, respectively.

Conclusions

The disagreement between PA-TD and EV®-CO measurements after anesthesia induction and after ICU admission, as well as the fact that thoracic bioimpedance did not adequately reflect pre- to postoperative changes in CO, questions the reliability of EV®-CO measurements in cardiac surgery patients and contrasts sharply with previous studies.
Keywords:Cardiac surgery  Hemodynamic monitoring  Non-invasive  Bioimpedance  Pulmonary artery catheter  Thermodilution