Identification of a recurrent pattern of false-positivity by Luminex HLA MHC class I single antigen bead testing |
| |
Affiliation: | Department of Pathology and Laboratory Medicine, Emory University, Atlanta, GA, United States |
| |
Abstract: | Previously, a distinct MHC class II Luminex-single antigen bead (SAB) pattern was described and attributed to antibodies targeting denatured antigens. In this study, we describe a distinct MHC class I reactivity pattern observed in 1.8% (105/5992) of samples resulted in 2017. The pattern displays reactivity to the following Luminex-SABs: HLA-A*33:03, A*36:01, A*80:01, B*54:01, B*53:01, C*06:02, C*07:02, C*18:02, C*14:02, C*03:03, C*03:04, and C*15:02. This pattern was identified in patients with no sensitization history, negative FlowPRA results, and antibody to self-antigen(s). Epitope analysis failed to reveal a common determinant(s) to explain this pattern of reactivity. Additionally, we found this pattern to be prevalent in female patients (62%) and also those with systemic lupus erythematosus (62%). Given these findings, we speculate this pattern likely represents false-positive reactivity, possibly due to antibody targeting denatured antigens or a specific peptide, molecular mimicry, autoimmunity, or a combination thereof. |
| |
Keywords: | HLA antibody MHC class I Luminex assay Single antigen bead |