Abstract: | PURPOSE: The present study was designed to examine the pharmacokinetics of a fluocinolone acetonide (FA) intravitreal implant in pigmented rabbits. METHODS: Pigmented rabbits were randomly assigned to receive either a 0.5 mg or 2.0 mg FA intravitreal implant (Retisert). Four animals were sacrificed per time point (2 hours; 2 weeks; and 3, 6, 9, and 12 months after implantation) for FA intraocular levels determination. RESULTS: In the vitreous, concentration of FA was relatively constant from the first time point, 2 hours, through 1 year, and dose-related, approximately seven- to eight-fold greater in the 2-mg implant. Concentrations of FA were generally higher in the vitreous (11-18 and 75-146 ng/g) and retina (42-87 and 224-489 ng/g) than in the aqueous humor (0.21-1.1 and 2.6-13.0 ng/g) for the 0.5- and 2-mg implants, respectively. Urine and plasma values were below the lower limit of quantitation (200 pg/mL) for all observations, indicating no evidence of systemic absorption. CONCLUSIONS: In this rabbit study, the Retisert provides relatively constant levels of FA in the posterior pole, which is consistent with previous reports of its clinical utility. |
| |