Indirect polarographic and cathodic stripping voltammetric determination of cefaclor as an alkaline degradation product
Authors:Rodrigues L N  Zanoni M V  Fogg A G
Institution:Departamento de Química Analítica, Instituto de Química, Araraquara SP, Brazil.
Abstract:Cefaclor is not reducible at a mercury electrode, but it can be determined polarographically and by cathodic stripping voltammetry as its initial alkaline degradation product which is obtained in high yield by hydrolysis of cefaclor in Britton-Robinson (B-R) buffer pH 10 at 50 degrees C for 30 min (reduction peak at pH 10, -0.70 V). Differential pulse polarographic calibration graphs are linear up to at least 1 x 10(-4) mol/l(-1). Recoveries of 93% of the cefaclor (n = 3) were obtained from urine spiked with 38.6 microg/ml(-1) using this polarographic method with 1 ml urine made up to 10 ml with pH 10 buffer. Using cathodic stripping voltammetry and accumulating at a hanging mercury drop electrode at - 0.2 V for 30 s, linear calibration graphs were obtained from 0.35 to 40 microg/ml(-1) cefaclor in B-R buffer pH 10. A relative standard deviation of 4.2% (eta = 5) was obtained, and the limit of detection was calculated to be 2.9 ng/ml(-1). Direct determination of cefaclor in human urine (1 ml of urine was made up to 10 ml with pH 10 buffer) spiked to 0.39 microg/ml(-1) was made (recovery 98.6%).