Ultrashort echo time imaging with bicomponent analysis
Authors:Du Jiang  Diaz Eric  Carl Michael  Bae Won  Chung Christine B  Bydder Graeme M
Affiliation:Department of Radiology, University of California, San Diego, California 92103-8226, USA. jiangdu@ucsd.edu
Abstract:
Biological tissues frequently contain different water compartments, and these often have distinct transverse relaxation times. Quantification of these may be problematic on clinical scanners because spin echo sequences usually have initial echo times that are too long to accurately quantify shorter relaxation time components. In this study, an ultrashort echo time pulse sequence was used together with bicomponent analysis to quantify both the short and long T(2) components in tissues of the musculoskeletal system. Feasibility studies were performed using numerical simulation, and on phantoms and in vitro tissues including bovine cortical bone, ligaments, menisci, tendons, and articular cartilage. The simulation and phantom studies demonstrated that this technique can quantify T(2) * and fractions of the short and long T(2) components. The tissues studies showed two distinct components with short T(2) *s ranging from 0.3 ms for bovine cortical bone to 2.1 ms for menisci, and long T(2) *s ranging from 2.9 ms for bovine cortical bone to 35.0 ms for articular cartilage. The short T(2) * fraction ranged from 18.5% for patella cartilage to 80.9% for ligaments. The results show that ultrashort echo time imaging with bicomponent analysis can quantify the short and long T(2) water components in vitro in musculoskeletal tissues.
Keywords:ultrashort TE  bicomponent analysis  short T2 component  long T2 component  T2*
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号