Laser ring catheters in excimer laser angioplasty |
| |
Authors: | S H Duda K R Karsch K K Haase P E Huppert C D Claussen |
| |
Affiliation: | Department of Radiological Diagnostics, Eberhard-Karls-University of Tübingen, Federal Republic of Germany. |
| |
Abstract: | Percutaneous excimer laser angioplasty is dependent on a reliable and effective energy transmission through fiberoptic devices. Compared with bare fibers, ring catheters provide improved flexibility and a central channel for insertion of guide wires. Peripheral laser catheters (7-9 F) were used in 32 patients with arterial occlusive disease in the lower limbs. The technical success rate was 91%. Vessel wall dissections occurred in nine cases but caused no clinical sequelae in the preliminary follow-up period (1-6 months). The clinical application of the catheter devices proved to be safe. |
| |
Keywords: | |
|
| 点击此处可从《Radiology》浏览原始摘要信息 |
|
点击此处可从《Radiology》下载全文 |