首页 | 本学科首页   官方微博 | 高级检索  
     


Serum metal ion concentration: comparison between small and large head metal-on-metal total hip arthroplasty
Authors:Hallows Rhett K  Pelt Christopher E  Erickson Jill A  Peters Christopher L
Affiliation:
  • ? Intermountain Orthopaedic Specialty Group, LDS Hospital, Salt Lake City, Utah
  • Department of Orthopaedics, University of Utah, University Orthopaedic Center, Salt Lake City, Utah
  • Abstract:
    The purpose of this study was to compare serum metal ion concentrations (chromium and cobalt) in 3 groups, 2 with metal-on-metal articulations and a control metal-on-polyethylene group. Forty-six patients with good to well-functioning hips were recruited for the study. Serum ion levels of all patients were drawn, and Harris Hip Score, University of California, Los Angeles activity score, and radiographs were performed. Serum chromium concentrations were significantly lower in the large head group compared with the small head group (P = .013). There was no difference in the cobalt concentrations between the 2 groups (P = .087). There was a significant difference between both metal-on-metal groups when compared with controls for both chromium and cobalt levels (P = .0005 and P = .0004, control vs small; P = .001 and P = .0001, control vs large, respectively).
    Keywords:serum metal ion concentration   total hip arthroplasty   chromium   cobalt