首页 | 本学科首页   官方微博 | 高级检索  
     


Simulation of time-resolved breast transillumination
Authors:E. B. de Haller  C. Depeursinge
Affiliation:(1) Laboratoire de Génie Médical, Ecole Polytechnique Fédérale de Lausanne, Champs-Courbes 1, 1024 Ecublens, Switzerland
Abstract:
A Monte Carlo simulation has been developed to predict the quality of time-resolved images of the breast by transillumination. The smallest diameter of a detectable carcinoma located in the breast has been computed. The simulation suggests that time-resolved imaging of the breast is possible and invaluable in the near infra-red (NIR) by transillumination. The enhancement of the transfer function by the introduction of time-resolved detection is limited by the contribution of noise at short integration times. The estimated diameter of the smallest detectable sphere is derived from the image quality index (IQI) theory and its value is around 4 mm. The simulated images of an absorbing sphere (approximating the carcinoma) within a homogeneous medium (approximating the surrounding tissue) show a significant improvement of the image with short integration time.
Keywords:Breast cancet  Image quality index  Monte Carlo  Time of flight  Transillumination
本文献已被 SpringerLink 等数据库收录!