首页 | 本学科首页   官方微博 | 高级检索  
     


The one-hour pad-weighing test: Reproducibility and the correlation between the test result,the start volume in the bladder,and the diuresis
Authors:Gunnar Lose,Jens Gammelgaard,Tine J. J  rgensen
Affiliation:Gunnar Lose,Jens Gammelgaard,Tine J. Jørgensen
Abstract:
To evaluate the reproducibility of the one-hour pad-weighing test as proposed by the ICS and to study the association between the test results, the initial volume in the bladder, and the diuresis, 18 females with stress or mixed incontinence underwent two separate tests. Bladder volumes were estimated by transabdominal ultrasonic scanning. A significant intra-individual variation in the test results was found. In about 50% of the patients the classification of the degree of leakage changed from one test to the other. This variation could be explained by variation in the fluid load on the bladder during the test, ie, the initial volume and the diuresis. For scientific purposes the one-hour pad-weighing test in its present design is not precise enough to allow reliable quantitation of urinary incontinence.
Keywords:one-hour pad-weighing test  reproducibility  diuresis  bladder volume
设为首页 | 免责声明 | 关于勤云 | 加入收藏