首页 | 本学科首页   官方微博 | 高级检索  
     


Local shielding of high dose rate brachytherapy in an operating theatre
Authors:Sephton R  Das K R  Coles J  Toye W  Pinder P
Affiliation:Physical Sciences Department, Peter MacCallum Cancer Institute, Melbourne, Australia. rsephton@petermac.unimelb.edu.au
Abstract:
A high-dose rate brachytherapy facility was installed into an established operating theatre by using local shielding in the form of mobile lead screens and by taking advantage of the ease with which staff movements can be controlled in an operating suite. This facility was inexpensive to develop, and has proved clinically efficient and entirely adequate from a radiation safety standpoint.
Keywords:
本文献已被 PubMed 等数据库收录!