Biomechanical modeling of acetabular component polyethylene stresses,fracture risk,and wear rate following press‐fit implantation
Authors:Kevin L. Ong  Steve Rundell  Imants Liepins  Ryan Laurent  David Markel  Steven M. Kurtz
Affiliation:1. Exponent Inc., 3401 Market St., Suite 300, Philadelphia, Pennsylvania 19104;2. Drexel University, Philadelphia, Pennsylvania;3. Stryker Orthopaedics, Mahwah, New Jersey;4. Providence Hospital, Southfield, Michigan
Abstract:
Press‐fit implantation may result in acetabular component deformation between the ischial‐ilial columns (“pinching”). The biomechanical and clinical consequences of liner pinching due to press‐fit implantation have not been well studied. We compared the effects of pinching on the polyethylene fracture risk, potential wear rate, and stresses for two different thickness liners using computational methods. Line‐to‐line (“no pinch”) reaming and 2 mm underreaming press fit (“pinch”) conditions were examined for Trident cups with X3? polyethylene liner wall thicknesses of 5.9 mm (36E) and 3.8 mm (40E). Press‐fit cup deformations were measured from a foam block configuration. A hybrid material model, calibrated to experimentally determined stress–strain behavior of sequentially annealed polyethylene, was applied to the computational model. Molecular chain stretch did not exceed the fracture threshold in any cases. Nominal shell pinch of 0.28 mm was estimated to increase the volumetric wear rate by 70% for both cups and peak contact stresses by 140 and 170% for the 5.9 and 3.8 mm‐thick liners, respectively. Although pinching increases liner stresses, polyethylene fracture is highly unlikely, and the volumetric wear rates are likely to be low compared to conventional polyethylene. © 2009 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 27:1467–1472, 2009
Keywords:THA  finite element  deformation  wear  polyethylene
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号