首页 | 本学科首页   官方微博 | 高级检索  
     

新型磁耦合式轴流血泵
引用本文:蔺嫦燕,李冰一,姜以岭,王景,陈力争,侯晓彤,孙衍庆. 新型磁耦合式轴流血泵[J]. 北京生物医学工程, 2001, 20(1): 69-72
作者姓名:蔺嫦燕  李冰一  姜以岭  王景  陈力争  侯晓彤  孙衍庆
作者单位:北京心肺血管疾病研究所;北京心肺血管疾病研究所;北京心肺血管疾病研究所;北京心肺血管疾病研究所;北京心肺血管疾病研究所;北京心肺血管疾病研究所;北京心肺血管疾病研究所
基金项目:国家科技攻关项目;969060214;
摘    要:为了更好地解决旋转式叶片血泵的密封问题,本文中采用磁耦合驱动方式。设计、加工一小型轴流血泵。血泵由泵体、直流无刷电机、叶轮、一对磁片、隔膜、导流槽和支撑架等部件构成。总长度125mm,容积147ml,材料是TC4钛合金,总重量380g,体外和在体试验中分别测试血泵的动力学输出、密封、产热和血液破坏等性能。结果显示:(1)体外试验中,血泵旋转速度8000rpm时输出为6.8L/min、120mmHg。(2)在120小时的连续动转过程中,血泵的表面温度变化较小;密封性能也较好。(3)24小时的在体试验中,血浆中游离血红蛋白含量逐渐增大,纤维蛋白未发生明显变化;除血泵的进出接口处和动物肾脏发现微小血栓外,脑部和其它腹部脏器未发现血栓。

关 键 词:轴流血泵动力学输出血液破坏密封产热
文章编号:1002-3208(2001)01-0069-03
修稿时间:1999-11-08

A New Magnetic Coupling Axial Flow Pump
LIN Changyan,LI Bingyi,JIANG Yiling,et al.. A New Magnetic Coupling Axial Flow Pump[J]. Beijing Biomedical Engineering, 2001, 20(1): 69-72
Authors:LIN Changyan  LI Bingyi  JIANG Yiling  et al.
Affiliation:LIN Changyan,LI Bingyi,JIANG Yiling,et al. Beijing Heart Hung & Blood Vessel Disease Institute,Beijing 100029
Abstract:For rotating blood pump, the sealing problem is a very important one to solve. In this paper, it is introduced that we designed and made a small axial flow pump, applying the magnetic coupling method. The pump consist of a pump body,a brushless DC motor, an impeller, a pair of magnetic discs, a spacer, a blood trough , a support frame, etc. The pump is made of titanium and is 125mm length, 147ml volume, total 380g of weight. Performances of outputting ,sealing, heat creating and damage to blood by the pump were investigated. Results showed for external experiment that: ①The flow rate of the pump was 6. 8L/min against 120mmHg pressure when the rotation speed was 8000rpm. ②The surface t6mperature of the pump didn't change obviously in 120 hours of rotation, and the sealing function was very well. ③For vivo experiment, free hemoglobin in blood plasma increased but fibrinogen didn't change obviously, and, at the end there was some micro thrombi at kidney of sheep and at inlet and outlet of the pump after 24 hours. There was no thrombus at the brain and other abdominal organs.
Keywords:Axial flow blood pump Dynamics output Damage to blood Sealing Heating
本文献已被 CNKI 维普 万方数据 等数据库收录!