首页 | 本学科首页   官方微博 | 高级检索  
     


Depth dose enhancement of electron beams subject to external uniform longitudinal magnetic fields: a Monte Carlo study
Authors:Earl M A  Ma L
Affiliation:University of Maryland School of Medicine, Baltimore 21201, USA. mearl001@umaryland.edu
Abstract:We studied the dose distributions from electron beams subjected to a longitudinal magnetic field applied to them before they reach the phantom. We found that specific combinations of the length and intensity of the magnetic field produced enhancement of the peaks of the central-axis depth-dose distributions. The EGS4 Monte Carlo system was used in this study. In the simulations, a uniform axial magnetic field parallel to the electron beam direction was applied to the air gap between the collimation and the phantom. We extensively studied the simplified case of an 18 MeV electron beam point source. Dose deposition was calculated for various magnetic field strengths, distances through which the magnetic field was applied, collimation sizes, and source to collimation distances. The magnetic field strengths varied from 0 to 3 T, the source-to-collimation distances studied were 50 and 95 cm, the collimation sizes studied were 10 x 10 and 20 x 20 cm2, and the distance through which the field was applied ranged from 10 to 20 cm. Specific combinations of these variables resulted in as much as a 70% enhancement of the peak dose relative to the surface dose. Finally, to determine how the geometry of a real accelerator affects the resulting dose distribution, we performed a full simulation of an Elekta SL20 linear accelerator and compared the results with the ideal case.
Keywords:
本文献已被 PubMed 等数据库收录!