3种仪器测量白内障患者水平角膜直径、平坦轴角膜曲率和陡峭轴角膜曲率的对比研究
引用本文:周歆,王从毅,李妍,折喜龙,党冠星. 3种仪器测量白内障患者水平角膜直径、平坦轴角膜曲率和陡峭轴角膜曲率的对比研究[J]. 延安大学学报(医学科学版), 2021, 19(4): 33-38. DOI: 10.19893/j.cnki.ydyxb.2020-0110
作者姓名:周歆  王从毅  李妍  折喜龙  党冠星
作者单位:西安市第四医院眼科,陕西 西安 710000
基金项目:西安市科技局科学计划项目[201805097YX53F31(6)]
摘    要:
目的 对比IOLMaster 700、OPD-ScanⅢ与MODi 2角膜地形图仪3种仪器在测量白内障患者水平角膜直径(white to white,WTW)、平坦轴角膜曲率(keratometry of flataxial,Kf)和陡峭轴角膜曲率(keratometry of steepaxial,Ks)的差异性和一...

关 键 词:水平角膜直径  平坦轴角膜曲率  陡峭轴角膜曲率  IOLMaster 700
收稿时间:2020-04-26

Comparative analysis of horizontal corneal diameter,flat axial corneal curvature and steep axial corneal curvature in cataract patients with three different instruments
ZHOU Xin,WANG Congyi,LI Yan,SHE Xilong,DANG Guanxing. Comparative analysis of horizontal corneal diameter,flat axial corneal curvature and steep axial corneal curvature in cataract patients with three different instruments[J]. Journal of Yanan University:Medical Science Edition, 2021, 19(4): 33-38. DOI: 10.19893/j.cnki.ydyxb.2020-0110
Authors:ZHOU Xin  WANG Congyi  LI Yan  SHE Xilong  DANG Guanxing
Affiliation:Cataract and Intraocular Lens Center,No.4 Hospital of Xi'an,Xi'an 710000,China
Abstract:
Objective To compare the difference and consistency of white to white(WTW),keratometry of flataxial(Kf) and keratometry of steepaxial(Ks) in cataract patients with IOL Master 700,OPD-Scan Ⅲ and MODi 2 corneal topographer.Methods A total of 54 patients (92 eyes) from Cataract and Intraocular Lens Center,No.4 Hospital of Xi'an during November to December 2019 were recruited.They were 17 males (28 eyes) and 37 females (64 eyes),with an average age of (52.83±15.00)years.Three instruments were used to measure WTW,Kf and Ks.One-way ANOVA analysis and LSD-t test were used to compare differences between instruments,and Pearson correlation analysis was used to analyze instrument correlation and Bland-Altman method to evaluate the consistency of instruments.Results When the WTW,Kf and Ks values were measured by three instruments,only the WTW values of IOLMaster 700 and OPD-scan Ⅲ,OPD-scan Ⅲ and MODi 2 corneal topographer showed statistically significant differences(P<0.001),while the differences of the remaining measured values were not statistically significant(P>0.05).In the correlation analysis,the measured values of the three instruments all showed significant positive correlation,with the correlation coefficient r between 0.75~0.98.When Bland-Altman method was used to evaluate instrument consistency,the measured WTW,Kf,and Ks values at 95% of the consistency threshold were respectively (5.43%,6.52%,4.34%),(4.34%,4.34%,and 7.61%),and (5.43%,5.43%,and 7.61%) of IOLMaster 700 and OPD-scan Ⅲ,IOLMaster 700 and MODi 2 corneal topographer,and OPD-scan Ⅲ and MODi 2 corneal topographer.Conclusion There was no difference in Kf and Ks values measured by the three instruments,and there was no statistically significant difference in WTW value measured by IOLMaster 700 and MODi 2 corneal topographer.However,the differences of Kf,Ks and WTW measured by the three instruments were clinically significant difference and could not be used interchangeably.
Keywords:Horizontal corneal diameter  Flat axial corneal curvature  Steep axial corneal curvature  IOLMaster 700  