Effect of a Stabilized Microbubble Contrast Agent on CW Ultrasound Induced Red Blood Cell Lysis In Vitro |
| |
Authors: | M. W. MILLER Ph.D M. AZADNIV Ph.D Y. DOIDA Sc.D A. A. BRAYMAN Ph.D |
| |
Affiliation: | Department of Biophysics, School of Medicine and Dentristry, The University of Rochester, Rochester, New York;Department of Biology, Shiga University of Medical Science, Seta, Otsu, Shiga, Japan |
| |
Abstract: | ![]() Human red blood cells (RBCs) in vitro at various fractional hematocrits (HCTs) were exposed for 60–120 seconds in a dialysis tubing vessel to 1 MHz continuous wave ultrasound (0–5 W/cm2 SPTA intensity); exposure vessels were either rotated at 200 rpm or stationary. Some RBC suspensions also contained Albunex® (ALX; a commercially-produced microbubble clinical ultrasound contrast agent) at final concentrations ranging from 0–41 μL/mL RBC suspension. Isonation was either by one transducer or by two opposing, continuously-gated, balanced transducers. For the vessel rotation / no rotation experiments, ultrasound-induced cell lysis was always increased with the ALX regimen relative to that of the no-ALX regimen at HCTs up to about 10%; at higher HCTs (including whole blood), ultrasound-induced hemol-ysis essectially ceased to occur. The data are consistent with reports of the ineffectiveness of continuous wave ultrasound at 3–5 W/cm2 to lyse cells in vitro at physiological cell densities with or without echocontrast medium supplemantation. |
| |
Keywords: | Albunex echo-contrast agents human red blood cells cell lysis |
|
|