摘    要:轴流式血泵转速过高、离心式血泵容易产生流动死区是造成血液损伤的重要原因,而混流式血泵能有效缓解轴流式血泵的转速过高以及离心式血泵的流动死区问题。基于此,本研究旨在探究闭式叶轮混流式血泵的性能效果。通过数值模拟的方法对闭式叶轮混流式血泵进行数值模拟,分析该类型血泵的流场特性及压力分布情况,探讨其水力性能以及可能对红细胞造成的损伤程度,并与半开式叶轮结构混流式血泵的数值模拟结果进行性能对比。结果表明:本研究中的闭式叶轮混流式血泵具有良好的性能,能够安全高效运行。该泵在5 L/min下能够达到100 mm Hg的扬程,血泵内流动均匀,没有明显的涡流、回流以及流动停滞现象,压力分布均匀合理,可有效地避免血栓;溶血指数平均值(HI)为4.99×10^-4,具有良好的血液相容性;与半开式叶轮混流式血泵相比,闭式叶轮混流式血泵扬程和效率更高、溶血指数平均值更小,且具有更好的水力性能及避免血液损伤的能力。通过本文研究结果,或能为闭式叶轮混流式血泵的性能评价提供依据。

关 键 词:混流式血泵  闭式叶轮  数值模拟  水力性能  溶血指数

Numerical simulation and performance analysis of mixed flow blood pump
LUO Jiping,HUANG Diangui,XU Bin.Numerical simulation and performance analysis of mixed flow blood pump[J].Journal of Biomedical Engineering,2020(2):296-303.
Authors:LUO Jiping  HUANG Diangui  XU Bin
Institution:(School of Energy and Power Engineering,University of Shanghai for Science and Technology,Shanghai 200093,P.R.China)
Abstract:The high rotational speed of the axial flow blood pump and flow separation of the centrifugal blood pump are the main causes for blood damage in blood pump. The mixed flow blood pump can effectively alleviate the high rotational speed and the flow separation. Based on this, the purpose of this study is to explore the performance of the mixed blood pump with a closed impeller. A mixed flow blood pump with closed impeller was studied by numerical simulation in this paper. The flow field characteristics and the pressure distribution of this type of blood pump were analyzed. The hydraulic performance of the blood pump and the possible damages to red blood cells were also discussed.At last, pump performance was compared with the mixed flow blood pump with semi-open impeller. The results show that the mixed flow blood pump with close impeller studied in this paper can operate safely and efficiently with a good performance. The pump can reach the pressure head of 100 mmHg at 5 L/min mass flow rate. Flow in the blood pump is uniform and no obvious separation or vortex occurs. Pressure distribution in and on the impeller is uniform and reasonable, which can effectively avoid the thrombosis of blood. The average mean value of hemolysis index is 4.99 × 10^-4. The pump has a good biocompatibility. Compared with the mixed flow blood pump with semi-open impeller, the mixed flow blood pump with closed impeller has higher head and efficiency, a smaller mean value of hemolysis index prediction, a better hydraulic performance and the ability to avoid blood damage. The results of this study may provide a basis for the performance evaluation of the closed impeller mixed flow blood pump.
Keywords:mixed flow blood pump  closed impeller  numerical simulation  hydraulic performance  hemolysis index