Transient micro-elastography: A novel non-invasive approach to measure liver stiffness in mice
Authors:Bastard Cécile  Bosisio Matteo R  Chabert Michèle  Kalopissis Athina D  Mahrouf-Yorgov Meriem  Gilgenkrantz Hélène  Mueller Sebastian  Sandrin Laurent
Affiliation:Echosens, Research and Development Department, Paris, France. cecile.bastard@echosens.com
Abstract:AIM:To develop and validate a transient micro-elastography device to measure liver stiffness(LS) in mice.METHODS:A novel transient micro-elastography(TME) device,dedicated to LS measurements in mice with a range of measurement from 1-170 kPa,was developed using an optimized vibration frequency of 300 Hz and a 2 mm piston.The novel probe was validated in a classical fibrosis model(CCl4) and in a transgenic murine model of systemic amyloidosis.RESULTS:TME could be successfully performed in control mice below ...
Keywords:Transient elastography  Micro-elastography  Liver stiffness  Liver  Mice  Amyloidosis  Fibrosis  Ultrasound  