首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of a pulse oximeter with an ear oximeter and an in-vitro oximeter
Authors:Nick Mackenzie MD
Affiliation:(1) Redwood Medical Clinic, 2900 Whipple Ave, 94062 Redwood City, CA
Abstract:
A pulse oximeter was compared with an ear oximeter for measurement of arterial hemoglobin saturation within the range of 70 to 100% in 11 healthy volunteer subjects. Two hundred seventy-seven pooled data points were obtained, and analysis was performed by means of linear regression. The accuracy was 3% (95% confidence limits). The pulse oximeter was also compared with an in-vitro oximeter, and accuracy was within 2%. The pulse oximeter was easy to use because, unlike the ear oximeter, it reqired no time-consuming instrument calibration or site preparation. In addition, the delays involved in taking an invasive sample, transporting it to the blood gas laboratory, and waiting for the results were eliminated. Saturation values were continually available, and placement and use ot the pulse oximeter sensor caused no discomfort to the volunteer subjects.
Keywords:Measurement techniques: pulse oximetry, ear oximetry, in-vitro oximetry  Monitoring: oxygen saturation  Oxygen: saturation  History: oximetry
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号