Displacement of Dental Implants in Trabecular Bone under a Static Lateral Load in Fresh Bovine Bone
Authors:Wilfried Engelke Prof. Dr. Dr  Alois Müller DDS  Oscar A. Decco DDS  María J. Rau Dipl. ‐Ing  Andrea C. Cura  Mara L. Ruscio  Michael Knösel Priv. ‐Doz. Dr.
Affiliation:1. Associate professor, Department of Maxillofacial Surgery, Georg‐August–University Hospital, G?ttingen, Germany;2. dentist and oral surgeon, private practice, G?ttingen, Germany;3. adjunct professor, Department of Bioengineering, Faculty of Engineering, Universidad Nacional de Entre Rios, Argentina;4. and Director, Dental Service, General Hospital Diamante, Entre Rios, Argentina;5. Department of Bioengineering, Faculty of Engineering, Universidad Nacional de Entre Rios, Argentina;6. research assistant, Faculty of Engineering, Universidad Nacional de Entre Ríos, Argentina;7. associate professor, Dentistry Center, Department of Orthodontics, Georg‐August–University Hospital, G?ttingen, Germany
Abstract:
Aim: The study aims to provide objective data for the displacement of titanium screw implants in trabecular bone specimens. One hundred Semados implants (Bego, Bremen, Germany) were inserted in bovine type IV bone specimens. All implants had a diameter of 3.75 mm; 50 implants had a length of 8.5 mm and 50 implants had a length of 15 mm. Insertion torque was determined at intervals of 10, 20, and 30 Ncm. Implants were loaded horizontally with 10, 20, and 30 N for 2 seconds. An indicator strip was attached to the implant abutment to allow direct observation of implant movement relative to the bone surface. Horizontal displacement was assessed with an accuracy of measurement of 10 µm. Seven implants got lost by visible loosening. Degree of displacement was subject to evaluation with all others. Those implants showed a mean displacement of 59 µm for 10 N (n = 100), 173 µm for 20 N (n = 99), and 211 µm for 30 N (n = 93). The mean displacement of 15‐mm implants (16, 37, 51 µm) was significantly lower compared with 8.5‐mm implants (103, 311, 396 µm) corresponding to 10, 20, and 30 N as lateral loads. Conclusions: Displacement of screw implants in trabecular bone can be detected and visualized using commercially available endoscopes with a high magnification. A lateral load of 20 N indicates a mean displacement of over 100 µm and therefore results in a critical displacement.
Keywords:contact endoscopy  displacement  implant length  insertion torque  lateral loading  primary stability
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号