Study on contact state of knee prosthesis with different specifications
Tian Qianying,Wu Xian,Fu Yanmian,Zhu Weiqiang,Xu Jifeng,Fang Ming.. Study on contact state of knee prosthesis with different specifications[J]. Orthopaedic Biomechanics Materials and Clinical Study, 2020, 17(4): 72-74
Authors:Tian Qianying  Wu Xian  Fu Yanmian  Zhu Weiqiang  Xu Jifeng  Fang Ming.
Abstract:Objective To analysis the contact state of the same model products with the same design but different specifications, and provide theoretical basis for the selection of wear test specifications. Methods Based on reasonable assumptions, ABAQUS was used to simulate the force of the knee joint system in the standing position, and the contact area test of different specifications was carried out by using Fuji prescale film (MW) for comparative analysis. Results Through simulation, with the increase of specifications, the contact area of the femoral-tibia increased, whose range was from 128 to 220 mm2, while the shape gradually widened. The contact stress decreased, whose range was from 20.73 to 32.70 MPa.Conclusion Comparing the comprehensive rateofchange of contact stress and contact area, as well as the market share of each specification, it is more reasonable to select specification of #4 or #5 for the knee wear test.
Keywords:Knee joint   Different specifications   Contact stress   Contact area   Wear test