两种固体模体水等效特性的比较研究
引用本文:王建华,王洵,任江平.两种固体模体水等效特性的比较研究[J].中华放射肿瘤学杂志,2011,20(3):236-239.
作者姓名:王建华  王洵  任江平
作者单位:1. 浙江省宁波市医疗中心李惠利医院放疗科,宁波,315041
2. 浙江省宁波市第二医院放疗技术科
3. 浙江省宁波市鄞州人民医院放疗科
摘    要:目的 对两种固体模体的水等效性及相关校准因子进行检测.方法 对美国MED-TEC公司虚拟水和德国PTW公司RW3两种同体模体不同深度处进行X线、电子线照射测量,得到深度电离量曲线与水模体中的进行比较,而水等效因子两种模体直接比较.按照国际原子能机构定义得到不同能量电子线深度比例因子(Cpl)与注量比例因子(hpl),比...

关 键 词:固体水模体  水等效

Dosimetric characteristics of water equivalent for two solid water phantoms
WANG Jian-hua,WANG Xun,REN Jiang-ping.Dosimetric characteristics of water equivalent for two solid water phantoms[J].Chinese Journal of Radiation Oncology,2011,20(3):236-239.
Authors:WANG Jian-hua  WANG Xun  REN Jiang-ping
Abstract:Objective To investigate the water equivalent of two solid water phantoms.Methods The X-ray and electron beam depth-ion curves were measured in water and two solid water phantoms,RW3 and Virtual Water.The water-equivalency correction factors for the two solid water phantoms were compared.We measured and calculated the range sealing factors and the fluence correction factors for the two solid water phantoms in the case of electron beams. Results The average differenee between the measurled ionization in solid water phantoms and water was 0.42%and 0.16%on 6 MV X-ray(t=-6.15.P=0.001and t=-1.65,P=0.419)and 0.21%and 0.31%on 10 MV X-ray(t=1.728,P=0.135 and t=-2.296,P=0.061),with 17.4%and 14.5%on 6 MeV electron beams(t=-1.37.P=0.208 and t=-1.47,P=0.179)and 7.0%and 6.0%on 15 MeV electron beams(t=-0.58.P=0.581 and t=-0.90,P=0.395).The water-equivalency correction factors for the two solid water Dhantoms varied slightly largely,F=58.54,P=0.000 on 6 MV X-ray,F=0.211.P=0.662 on 10 MV X.ray,F=0.97.P=0.353 on 6 MeV electron beams,F=0.14,P=0.717 on 15 MeV electron beams.However,they were almost equal to 1 near the referenee depths.The two solid water phantoms showed a similar tread of Cpl increasing(F=26.40,P=0.014)and hpl decreasing(F=7.45,P=0.072)with increasing energy.Conclusion The solid water phantom should undergo a quality control test before being clinical use.
Keywords:SoIid Water phantom  Water equivalent