Assessment of colour doppler tissue imaging using test-phantoms
Authors:A.D. Fleming   W.N. McDicken   G.R. Sutherland  P.R. Hoskins
Affiliation:

Department of Medical Physics, Royal Infirmary of Edinburgh, Lauriston Place, Edinburgh EH3 9YW, Scotland, UK

Department of Cardiology, Western General Hospital, Edinburgh EH4 2XU, Scotland, UK

Abstract:
An investigation has been carried out on the velocity resolution, spatial resolution and accuracy of Doppler images as part of a study into the Doppler display of cardiac tissue motion. Test-phantoms were designed to perform this work and images were captured on a computer. The characteristics of the phantom images and of the image capture process were studied. The smallest spatial detail that was observed in the Doppler image was 3 mm by 3 mm. Doppler receive gain and Doppler ensemble size both affected velocity resolution. Different target materials gave different measures for velocity resolution. This could be related to the different back-scatter intensities of the materials.
Keywords:Doppler imaging   Velocity resolution   Spatial resolution   Accuracy   Echocardiography   Ultrasound