首页 | 本学科首页   官方微博 | 高级检索  
     


Source of the slippage in the universal joints of the Hoffmann external fixator
Authors:F. L. I. P. Drijber  J. B. Finlay  T. K. Moroz  C. H. Rorabeck
Affiliation:(1) Orthopaedic Research Laboratory, University Hospital, University of Western Ontario, Terminal ‘A’, PO Box 5339, N6A 5A5 London, Ontario, Canada;(2) Division of Orthopaedics, University Hospital, Terminal ‘A’, PO Box 5339, N6A 5A5 London, Ontario, Canada
Abstract:
An investigation was conducted to determine what improvements in the resistance to slippage could be obtained in selected interfaces (rod/clip torsional, cheek/bowl and cheek/clip) of the Hoffmann external fixator. The modification involved changing the standard wing-nut clamp for a bolt with a thread of 7 mm and a 1 mm pitch and placing an FAG 28–303 thrust-bearing (needle roller and cage assembly) between the bolt and the cheek. The results showed a significant improvement in the slippage values of all interfaces; increases of approximately six times were obtained at all torque values of the wing-nut clamp or fastener tested. Such improvements would markedly increase the reliability of external fixation systems and thus reduce the incidence of loss-of-reduction of fracture due to slippage of the universal joint.
Keywords:External fracture fixation  Slippage  Universal joint
本文献已被 SpringerLink 等数据库收录!