介入超声印压系统对活体舒张心肌黏弹性的检测作用
引用本文:李进嵩,张春晓,邓辉胜,袁乔英,江永红,朱悫,黄晶. 介入超声印压系统对活体舒张心肌黏弹性的检测作用[J]. 解放军医学杂志, 2017, 42(8). DOI: 10.11855/j.issn.0577-7402.2017.08.04
作者姓名:李进嵩  张春晓  邓辉胜  袁乔英  江永红  朱悫  黄晶
作者单位:1. 610072,成都 四川省医学科学院、四川省人民医院心内科;2. 400074,重庆 重庆交通大学土木建筑学院;3. 400010,重庆 重庆医科大学附属第二医院心内科
基金项目:国家自然科学基金,四川省卫生厅资助项目(110210)This work was supported by the National Natural Science Foundation of China,the Foundation of Health and Family Planning Commission of Sichuan Province
摘    要:
目的 探讨介入超声印压系统(IUIS)对活体心脏舒张期舒张心肌黏弹性的检测作用.方法 应用IUIS以印压法对陈旧性心肌梗死模型组(OMI组,n=8)和假手术组(n=8)犬在舒张期进行印压蠕变实验,以三参量固体黏弹性模型拟合实验数据,提取舒张心肌瞬时弹性模量(E1)、松弛模量(E∞)、蠕变弹性模量(E2)、黏性阻尼系数(η)等黏弹性参数,并进行比较.结果 OMI组E1、E∞、E2、η参数(分别为27.81±6.74kPa、17.87±3.59kPa、49.54±14.35kPa、1.97±0.78Pa.s)与假手术组(分别为6.78±2.43kPa、4.52±1.56kPa、16.82±12.37kPa、0.66±0.40Pa.s)比较均明显增加,差异有统计学意义(P<0.05).结论 IUIS系统及其印压蠕变实验方法 可用于活体舒张心肌黏弹性的评估.三参量固体黏弹性模型可用于描述舒张心肌蠕变特性.梗死心肌弹性模量和黏性阻力增加.

关 键 词:介入超声印压系统  舒张心肌  印压蠕变实验  三参量黏弹性模型  黏弹性参数

Detection of the viscoelasticity of relaxed myocardium in vivo by intervention ultrasound indentation system
LI Jin-song,ZHANG Chun-xiao,DENG Hui-sheng,YUAN Qiao-ying,JIANG Yong-hong,ZHU Que,HUANG Jing. Detection of the viscoelasticity of relaxed myocardium in vivo by intervention ultrasound indentation system[J]. Medical Journal of Chinese People's Liberation Army, 2017, 42(8). DOI: 10.11855/j.issn.0577-7402.2017.08.04
Authors:LI Jin-song  ZHANG Chun-xiao  DENG Hui-sheng  YUAN Qiao-ying  JIANG Yong-hong  ZHU Que  HUANG Jing
Abstract:
Objective To explore the viscoelasticity of relaxed myocardium in vivo through indentation method with an intervention ultrasound indentation system (IUIS). Methods Old myocardial infarction (OMI) models of canine were established by ligating left anterior descending branch of coronary artery for 3 months. The indentation creep tests were used respectively in OMI group and shame group (n=8, each) by IUIS in middle and advanced diastole stage in vivo. Test data were processed with three-parameter solid viscoelasticity model, and the viscoelastic parameters, such as instantaneous elastic modulus (E1), relaxation modulus (E∞), creep elastic modulus (E2) and viscous damping coefficient (η) in normal and infarcted myocardium were obtained and compared. Results All the parameters of E1, E∞, E2 and η increased obviously in OMI group than in sham group shown as follows: 27.81±6.74kPa vs. 6.78±2.43kPa; 17.87±3.59kPa vs. 4.52±1.56kPa; 49.54±14.35kPa vs. 16.82±12.37kPa and 1.97±0.78Pa.s vs. 0.66±0.40Pa.s. The differences were statistically significant (P<0.05). Conclusions IUIS is a feasible method to assess the viscoelasticity of relaxed myocardium in vivo. Three-parameter viscoelasticity model can be used to describe creep properties of relaxed myocardium. Both elastic modulus and viscosity resistance have increased in infarcted myocardium.
Keywords:intervention ultrasound indentation system  relaxed myocardium  indentation creep test  three-parameter viscoelastic model  viscoelastic parameters