The plantaris tendon graft: An ultrasound study |
| |
Authors: | CPT Stephen L. Simpson MD MC Michael S. Hertzog MD MC COL Roberto H. Barja MD MC |
| |
Affiliation: | Orthopaedic Surgery Service, Eisenhower Army Medical Center, Fort Gordon, GA 30905-5650. |
| |
Abstract: | ![]() To assess the value of ultrasound in imaging the plantaris tendon, ultrasounds were done on 26 legs in 25 patients who had a variety of surgical procedures on the posterior leg. Eighteen ultrasounds were done and interpreted before operation and eight after operation by a radiologist. The sonographic results were compared with the status of the plantaris as determined at surgery. The plantaris was present in 21 of 23 patients for a prevalence of 91%. Eighteen plantaris tendons were well visualized on ultrasound for a sensitivity of 86%. Of the three false-negative results, two occurred with thinned plantaris tendons that were less than 2 mm thick and hence considered inadequate for grafting. Therefore, the sensitivity for detecting a tendon suitable for grafting was 95%. All patients with positive studies had a plantaris tendon present for a specificity of 100%. A learning curve for the radiologist was demonstrated. Ultrasound appears to be effective in the preoperative imaging of the plantaris tendon. |
| |