首页 | 本学科首页   官方微博 | 高级检索  
检索        

角膜厚度对两种眼压测量方法的影响
引用本文:鲍捷,代艳,陈晓明.角膜厚度对两种眼压测量方法的影响[J].中国实用眼科杂志,2005,23(3):272-274.
作者姓名:鲍捷  代艳  陈晓明
作者单位:610041,成都,四川大学华西医院眼科
摘    要:目的 比较非接触眼压计 (non -contacttonometer ,NCT)和Goldmann压平眼压计测量眼压的差异 ,并分别探讨中央角膜厚度 (centralcornealthickniss ,CCT)对这两种测量方法的影响。方法 对 1 0 8例拟接受PRK或LASIK手术的患者行CCT ,NCT和Goldmann压平计眼压测量。结果NCT和Goldmann压平眼压计测得的眼压均值具有显著性差异 (F =89 .70 4 4,P <0 . 0 1 )。CCT与NCT和Goldmann压平眼压计测量值呈正相关 ,相关系数分别是r =0. 4 96 0 (t =8 .356 3,P <0 .0 0 1 )和r =0 . 2 1 1 3(t =3. 1 6 2 3,P <0 .0 0 1 )。结论 NCT和Goldmann压平眼压计测量眼压值有差异 ,NCT测量值大于Goldmann压平眼压计 ,CCT对NCT的影响大于Goldmann压平眼压计。

关 键 词:眼压计  NCT  眼压测量  CCT  角膜厚度  影响  LASIK手术  相关系数
修稿时间:2004年10月1日

The influence of central corneal thickness on the measurements of intraocular pressure with two kinds of tonometer
BAO Jie,CHEN Xiao ming.The influence of central corneal thickness on the measurements of intraocular pressure with two kinds of tonometer[J].Chinese Journal of Practical Ophthalmology,2005,23(3):272-274.
Authors:BAO Jie  CHEN Xiao ming
Institution:BAO Jie,CHEN Xiao ming. Department of Ophthalmology,West China Hospital,Sichuan University,Chengdu 610041,China
Abstract:Objective To compare the difference of intraocular pressure measured by non contact tonometer (NCT) and Goldmann applanation tonometer and to research the influence of central corneal thickness on the measurements of IOP obtained with NCT and Goldmann applanation tonometer.Methods central corneal thickness,IOP measured by NCT and Goldmann applanation tonometer were measured on 108(216 eyes)patients before PRK or LASIK.Results There is significant difference between IOP measured by NCT and IOP by Goldmann applanation tonometer (F=89 7044, P <0 01).CCT was positive relative to the IOP measured by NCT and Goldmann applanation tonometer.The correlated coefficients are r=0 4960(t=8 3563, P <0 001) and r=0 2113(t=3 1623, P <0 001) respectively.Conclusions There is difference between IOP measured by NCT and IOP by Goldmann applanation tonometer.IOP measured by NCT was higher than which measured by Goldmann applanation tonometer.The influence of central corneal thickness on the measurements of intraocular pressure with NCT was bigger than Goldmann applanation tonometer.
Keywords:Central corneal thickness  Non-contact tonometer  Goldmann applanation tonometer
本文献已被 CNKI 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号